Description
For additional information
ask a questionOD Plate is the measurement verification instrument for microplate photometer MPP-96 HiPo. The instrument is designed to verify the accuracy and precision of measurements of the photometer at 6 levels of nominal optical density: 0.3; 0.6; 1.0; 2.0; 3.0; 4.0 OD. The standard range is at following 17 Centre Wavelengths (CWL (± 1 nm)): 405, 414, 450, 480, 492, 515, 540, 550, 560, 568, 580, 594, 620, 632, 650, 690, 700.
Optionally, the plate can be verified at any following Centre Wave-lengths (CWL (± 1 nm)): 400, 405, 414, 420, 422, 430, 436, 440, 442, 445, 450, 455, 458, 460, 470, 473, 480, 486, 488, 492, 500, 505, 508, 510, 515, 520, 532, 540, 546, 550, 560, 568, 580, 589, 594, 600, 610, 620, 632, 636, 640, 647, 650, 656, 660, 671, 676, 680, 685, 690, 694, 700 nm.
Instrument is provided in a shockproof case and USB drive containing the verification data and software.
Included is copy of the calibration certificate of reference materials issued by accredited bodies.
-
Microplate Photometer HiPo MPP-96 is a compact tabletop device for measuring the results of ELISA and microbiological studies in 96-well microplates. Photometer is controlled and outputs data via computer. The …
read more