Description

For additional information

ask a question

OD Plate is the measurement verification instrument for microplate photometer MPP-96 HiPo. The instrument is designed to verify the accuracy and precision of measurements of the photometer at 6 levels of nominal optical density: 0.3; 0.6; 1.0; 2.0; 3.0; 4.0 OD. The standard range is at following 17 Centre Wavelengths (CWL (± 1 nm)): 405, 414, 450, 480, 492, 515, 540, 550, 560, 568, 580, 594, 620, 632, 650, 690, 700.
Optionally, the plate can be verified at any following Centre Wave-lengths (CWL (± 1 nm)): 400, 405, 414, 420, 422, 430, 436, 440, 442, 445, 450, 455, 458, 460, 470, 473, 480, 486, 488, 492, 500, 505, 508, 510, 515, 520, 532, 540, 546, 550, 560, 568, 580, 589, 594, 600, 610, 620, 632, 636, 640, 647, 650, 656, 660, 671, 676, 680, 685, 690, 694, 700 nm.

Instrument is provided in a shockproof case and USB drive containing the verification data and software.
Included is copy of the calibration certificate of reference materials issued by accredited bodies.

Specification

Nominal optical density levels 0.3; 0.6; 1.0; 2.0; 3.0; 4.0 OD (±0.1 OD)
Available verification wavelengths range 400 - 700 nm
Overall dimensions (W×D×H) 86 x 128 x 12 mm
Weight 0.3 kg

Send inquiry to Biosan

Software download

Please provide serial number to start download

Company Company