Description
For additional information
ask a questionLine-Gene 9600 is product of Bioer’s Real-Time PCR detection system family. Based on Line-Gene 9600’s large sample capacity, wide temperature range and multi-channels, Line-Gene 9600 provides brand-new & humanized running and operation interface for better operating experience.
Line-Gene 9600 adopts Ferrotec specially-made Peltier, advanced fibber optic technology, new global wide-range power supply, the unique patented block dissipation technology and bottom detection mode, which provides better heating/cooling rate, temperature accuracy, uniformity and stability. Line-Gene 9600 has up to 96 sample capacity and fast speed of dualcolor scanning, and the design of 4 channel detection also covers the whole fluorescence detection wavelength.
Newly-designed automatic hot-lid without manual operation ensures constant pressure for various tube heights. This auto hot-lid mode could be easily integrated into various automatic systems.
Newly-improved Line-Gene 9600 Real-Time PCR detection system can satisfy various clients’ requirement either for research or clinical use.
FLUORESCENCE CHANNELS LIST
- F1: FAM, SYBR Green
- F2: VIC, HEX, TET, JOE, CY3, TAMRA
- F3: ROX, TEXAS RED
- F4: CY5, Quasar-670
SOFTWARE FEATURES
- Absolute Quantification
- Relative Quantification
- SNP Analysis
- Data Automatic Analysis
- Melting Curve Genotyping
- Gradient
- HRM Multi-channel
- Crosstalk Correction
- Background Correction
- Automatic Gain
- Customized Parameters
Sample capacity | 0.2ml single tube (transparent cap), 96×0.2(0.1)ml Plate (transparent cap), 8-strip tubes (transparent cap) |
---|---|
Dynamic range | 1~10^10 copies |
Excitation wavelength | 300-800 nm |
Emission wavelength | 500-800 nm |
Channels | 4 |
Temperature setting range | 4 ~99.9°C (increment: 0.1°C) |
Heating/coolling rate | 4.0 °C/s |
Temperature accuracy | ≤±0.1°C |
Temperature fluctuation | ≤±0.1°C |
Temperature uniformity | ≤±0.3°C |
Temperature control mode | Block/6-zone gradient |
Sample volume | 1 ~ 100 μl |
Gradient temperature range | 1 ~ 30°C |
Hot-lid temperature range | 30 ~110°C (adjustable default: 105°C, Automatic Hot-lid) |
Repeatability | 5% |
Scan mode | Full plate simultaneously |
Program | Max. 20 segments for each program, max. 99 cycles |
Continuous operation | + |
Scan period | 5.5 s |
Operating system | Windows 7/8/10/11 |
Software required | Microsoft Excel |
Nominal operating voltage | 100-240V; ~50/60Hz |
Power consumption | 600W |
Dimensions (W×D×H) | 386×410×352 mm |
Number of sockets | USB adapter, Bluetooth adapter |
Certification | Ferrotec pelitier / CE / Met / RoHS2 / PICC product quality liability insurance |